Engelsk titel: Intraprosthetic luxation of trapeziometacarpal joint prosthesis Läs online Författare: Schulz, Anne Lysgaard ; Kirkeby, Lone ; Baek Hansen, Torben Språk: Dan Antal referenser: 5 Dokumenttyp: Fallbeskrivning UI-nummer: 20030119

Tidskrift

Ugeskrift for Laeger 2019;181(23)2237-8 ISSN 0041-5782 E-ISSN 1603-6824 KIBs bestånd av denna tidskrift Denna tidskrift är expertgranskad (Peer-Reviewed)

Sammanfattning

Total joint replacement is used for treating osteoarthritis of the trapeziometacarpal joint. The prosthesis is a ball and socket design resembling a total hip prosthesis and with the same risk of head luxation. To reduce this risk, dual mobility articulations have also been introduced in prosthesis of the thumb, and this is a case report of luxation of the polyethylene liner in a dual mobility prosthesis, illustrating that introducing new designs in prosthesis for the basal joint of the thumb may lead to similar complications as are seen in the hip joint.